- MPS Robotic Pro:
- Go to Extended User Level/Options/Shutdown
- Turn power off to the unit
- MPS -xt or MPS Robotic:
- Remove the syringe or tool from the MPS using the hand held keypad
- Turn off power
- Manually position the injection unit over a wash station or tray holder so that the unit has somewhere safe to rest while shut down
- TDU/TDU 2/TD 3.5+/TDS/Pyrolysis
- Replace any absorbent packed tube or liner with an empty glass equivalent.
- DHS – No additional actions required
- UPC+
- Disconnect coolant feed line (blue) from the TDU and place end in a beaker
- Press the “Pump” button on the back of the UPC+ until all coolant has drained from unit
- Reconnect coolant feed line
- Power off UPC
- C506 – Turn off power
- ODP
- Shut off gases connected to ODP pneumatic box if possible to stop make up and humidifier flow
- If heated by C200 or other heating source, turn power off to that unit
- If heated by the GC, no additional actions required
GERSTEL service representatives will continue to be available by either calling 410-204-7259 or via email at service@gerstelus.com Please contact us if you have any questions.